The Insitec ALISS is an at-line particle size analyzer for dry processes
More information:
 
Overview of the Insitec ALISS At-Line Particle Size Analyzer Overview
 
Accessories for Insitec ALISS At-Line Particle Size Analyzer Accessories
 
Systems

You are here:HomeProcess productsLaser diffractionSystemsInsitec ALISS specification

Insitec ALISS

At-line particle size analyzer for dry processes.

  Ideal for use when centrally located among multiple mills
 
  Rapid particle size analysis, based on laser diffraction, for particles in the size range 0.1 to 1000 µm
 
  Fully automated analytical procedure, no sample preparation required
 
  Easily upgraded for on-line applications
 

Award winning technology
Winner of the 2005 Powder/Bulk Solids Innovation Award (processing category).

Designed to meet industrial requirements for at-line particle size analysis, Insitec ALISS is particularly useful for process optimization, commissioning and on plant QC applications.

The capability to easily upgrade to meet a requirement for on-line analysis provides capital investment protection.

The Insitec ALISS is designed to meet industrial requirements for at-line particle size analysis
The Insitec ALISS is designed to meet industrial requirements for at-line particle size analysis
Laser Diffraction
Particle size analysis is performed using industry standard laser diffraction technology.


Process Interface
Large sample volumes can be analyzed easily and no sample preparation is required.


Automation & control
The automation of the analyzer system and connection to the plant control system is facilitated via Malvern Link, a flexible approach accepting custom user interfaces, and a variety of PLC communication protocols


Solutions
Insitec ALISS is suitable for at-line particle size analysis in a broad range of different process streams, from sectors as diverse as cement and powder coatings.
 
Overview of the Insitec ALISS At-Line Particle Size Analyzer Specifications for Insitec ALISS At-Line Particle Size Analyzer Accessories for Insitec ALISS At-Line Particle Size Analyzer

Summary

The configuration can be specified to suit particular applications, however the base system is configured as follows:

General

Intrinsic safety None
Analyzer
  • IP66 rated optical sensor comprising solid state laser diode, multi-element detector, high speed parallel processor, process sample flow cell, communications interface and interconnecting cables (150m). Range of fixed optical arrangements available, allowing different size ranges to be measured
  • Flow cell suitable for inline or bypass eductor configurations, with range of path lengths for compatibility with process lines and particulate loading. The flow cell comprises the central region of the instrument containing the sample measurement zone.
  • Triclamp assembly of optical components for simplified maintenance.
  • The cell windows are purged with dry clean air to protect them from contamination.
  • The flow cell is made of 316L stainless steel. Sample measurement zone and flow path are of seamless construction to comply with cGMP requirements. All seals are rated to IP66.
  • Process connection via tri-clamp fittings to BS4825
Installation
Static unit - Fit to pipe or frame close by to sampling points
Process Interface Kit

Standard on every system:

  • Dispersion provided by 90° Stainless Steel Venturi Eductor
    (alternate dispersion systems available as options)
  • 80cm Stainless Steel test flute with pre-drilled holes to allow rapid optimization of probe configuration.
  • 2 meters flexible inlet pipe
  • 2 meters flexible outlet pipe
  • 3 fixed sampling point flanges + 80 cm standard flutes

Standard system only:

  • Inlet and outlet ball valves with triclamp fittings (Manual)

Automation option:

  • Pneumatic inlet and outlet pinch valves replace the manual ball valves
Air supply

Air Panel Assembly

  • Three adjustable air flow meters (lockable to prevent inadvertent adjustment)
  • Low-flow purge to protect windows when system is idle
  • Cartridge filters

Automation option only:

Air Cabinet

  • Automated air panel housed in a lockable IP56 polyester cabinet.
  • Three adjustable air flow meters, automatic valves and flow sensors.
  • Boost air facility
  • Pre-equipped with 4/2 air distributors plus pressure sensors enabling automatic operation of up to six pneumatic pinch valves
  • All automatic functions configurable via Malvern Link
     
Electronic enclosure
  • Lockable IP66 rated cabinet - housing:
  • Communication interface PCB - allows RS232 or RS422 connection to the PC.
  • Maximum cable length is 15m using RS232. The RS422 connection should be used for longer length requirements.
  • Diagnostic LED's at the Optical sensor
  • Voltage is auto sensed in range 100 to 240 volts

Automation

Automation cabinet

Option:

  • Hardware supporting full Malvern Link functionality housed in a lockable IP56 polyester cabinet, containing
  • Wago, analog and digital output modules
  • Power Supply
  • Switches for installation testing
  • Terminals for field wiring
  • Click here for more details
Software Suite

The following software is pre-installed as standard with each process system (and also on CD-ROM). They are compatible with Microsoft Windows NT4.0 and 2000 for use on Intel compatible PC and available in English, French, German, Spanish and Japanese languages.

  • RTSizer for instrument configuration, data acquisition and particle size analysis; includes high concentration software module, patented technique to enable measurement under high or fluctuating process loadings (5-95% transmission)
  • Malvern Link for instrument control and operator interface, including license for award winning Intellution iFix SCADA software.
  • PCAnywhere for remote diagnostics. (Requires access to a telephone line)
Malvern Link Software and Computer

Standard system only:

  • Computer excluded. Malvern Process Systems can supply IP54 laptop on request.

Automation options:

  • Malvern Link options such as Plant Optimizer, Batch Tracker...Access to Engineering Mode of Malvern Link.
  • "Toughnote" IP54 laptop computer

Miscellaneous

Installation & Training
  • Includes two days training at Malvern office plus three days on-site training. Upon completion of training, Malvern Process Systems will offer on-site applications assistance at a 10% discount to our standard daily rate for the first twelve months. T&E charged at cost.
  • Installation, commissioning and additional training days excluded. User is responsible for installation, commissioning and application development.
Standard documentation

The following documents are supplied on CD-ROM in PDF format and available in English, French, German, Spanish and Japanese languages.

  • Installation manual
  • User manuals
Site requirements
  • Location for instrument to be specified (following consultation from Process Specialist engineers):
  • On-line (eductor) configuration: Sample ports (supplied by Malvern Process Systems) to be connected to process line
  • In-line (flowcell) configuration: Process line broken and flanges added to connect to instrument.
  • A dry clean air or nitrogen supply of a maximum of 50m3/hour at 4 bar g (18scfm at 60 psi g) is required at instrument location to provide for window purge and motive air.
  • A fused spur rated at 100-240 Volts AC, 5 AMPS and ground connection is required on site to provide power for sensor and interface enclosure.
  • Local electricity supply
  • Instrument - PC data communications (cable supplied with instrument) to be installed from instrument location to PC location
  • PC to be installed in control room type environment free from particulates, and controlled temperature/humidity.

Specification

Lens size range

Choice of four size ranges (one lens is delivered as standard, others can be ordered as options):

Lens focal length (mm)

Size range (µm)

100

0.1 - 200

200

1.0 - 400

300

1.5 - 600

450

2.25 - 1000

Intrinsically safe
No
Materials
  • 316L for all metallic parts, with high grade finish
  • Glass windows (for measurement zone)
  • Viton O-rings and EPDM triclamp gaskets (other seal materials can be supplied)
Fittings

Triclamped for simplified cleaning

Transmission

5 - 98 %

Accuracy ±2% on Dv(50) reported using the verification reticle
Air requirements Dry, oil-free air containing no particulates
6 bar g (90 psi g) with a flowrate of 30 m³/h (1000 ft³/h) required as motive gas for eductor and for window purge

Note: other gases (e.g. nitrogen) can be used instead of air as appropriate
Maximum operating pressure 10 bar (g)
Operating temperature (ambient -10°C to +70°C
Industrial protection Dust-tight and waterproof: Optical head IP66, cabinets IP65
Maximum distance from instrument to PC 500 m
Software RTSizer (for instrument control)
Malvern Link (for system automation and data link)

All software runs on MS-Windows 2000 or above
Regulatory compliance Validation and 21 CFR part 11options
Automation

Options:

  • PLC in cabinet
  • 2-off analogue outputs (optional 4-off analogue inputs for logging extra process parameters)
  • valve sequencing for automatic background, sample path cleaning, start / stop
  • 8-off digital I/O (for valve switching and alarms)
  • OPC server to plant DCS
Dimensions 1000mm width, 1620mm height, 800mm depth (Automated)
1000mm width, 1240mm height, 800mm depth (Manual)
Weight (approx) 150kg (Automated), 175kg (Manual)