Sample handling and dispersion is simplified by the use of standard 600 to 1000 ml laboratory beakers in place of a fixed tank. Its robust construction and relatively small size have made the Micro very popular for offshore and mineral exploration applications.
|
Single Measurement range.
Measures materials from 0.3 µm to 300 µm.
|
|
Dispersion Type.
Wet, Capacity 600 to 1000 ml
|
|
Dispersion Mechanism.
Dip-in pump and stir head with continuously variable single shaft pump and stirrer and continuously variable ultrasonic probe. The spring counterbalanced stir head is lowered into a standard 600 ml to 1000 ml laboratory beaker in which the sample is dispersed.
|
|
Dispersion control Operation.
Manual control via splash-resistant touch pad on unit.
|
|
User friendly Software.
Configurable for one-button operation. Standard operating routines can be set up to reduce operator error.
|
|
Accuracy and Reproducibility.
Accuracy: ± 3% on the Dv50 using Malvern Quality Audit Standards Instrument-to-instrument reproducibility: Better than 3% RSD on the Dv50 of the Malvern Quality Audit Standard.
|